To the editor
In reference to a recent publication by Flisiak et al,1 we would like to draw attention to the quickly evolving approach to rapid point-of-care antibody cassette tests for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Numerous manufacturers offer rapid point-of-care cassette tests, which detect immunoglobulin M (IgM) and immunoglobulin G (IgG) antibodies to SARS-CoV-2 (Table 1).
Test | Manufacturer | Sensitivity | Specificity | Total accuracy | Recommended sample type | Sample size | Time |
---|---|---|---|---|---|---|---|
ACCU-TELL COVID-19 IgG / IgM Cassette | AccuBioTech Co., Ltd., China | IgG: 97.4%; IgM: 86.8% | IgG: 99.3%; IgM: 98.6% | IgG: 98.9%; IgM: 96.1% | serum, plasma, whole blood | 10 μl | 10 min |
2019-nCoV IgG / IgM Rapid Test Cassette (Whole Blood / Serum / Plasma) | Hangzhou AllTest Biotech Co., Ltd., China | IgG: 100%; IgM: 85% | IgG: 98%; IgM: 96% | IgG: 98.6%; IgM: 92.9% | serum, plasma, whole blood | 10 μl of serum or plasma; 20 μl of fingertip blood or whole blood | 10 min |
Rapid test 2019-nCOV blood | myLAB, Poland | – | – | – | whole blood | 20 μl | 10 min |
The StrongStep COVID-19 IgG / IgM Combo Test | Liming Bio-Products Co., Ltd., China | IgG: 93.1%; IgM: 64.7% | IgG: 100%; IgM: 100% | – | serum, plasma, whole blood | 10 μl | 15 min |
Novel Coronavirus (2019-nCoV) Antibody IgG / IgM Assay (Colloidal Gold) | Avioq Biotechnology Co., Ltd., China | – | – | – | serum, plasma, whole blood | 10 μl | 15 min |
2019-nCoV IgG & IgM Antibody Determination Kit | Beijing Diagreat Biotechnologies Co., Ltd., China | – | – | – | serum, plasma, whole blood | – | 15 min |
COVID-19 IgM / IgG Rapid Test | BioMedomics, Inc., United States | 88.66% | 90.63% | – | serum, plasma, fingertip blood, whole blood | – | 15 min |
OnSite COVID-19 IgG / IgM Rapid Test | CTK Biotech, Inc., United States | 96.9% | 99.4% | – | – | – | 10 min |
STANDARD Q COVID-19 IgM / IgG Duo | SD BIOSENSOR, Inc., Korea | 81.8% | 96.6% | – | serum, plasma, whole blood | 10 μl | 10–15 min |
2019-Novel Coronavirus (2019-nCoV) IgG / IgM GICA Rapid Test Kit | Shenzhen Bioeasy Biotechnology Co., Ltd., China | 89.56% (plasma: 90.3%; serum: 89.47%; whole blood: 88.89%) | 99.8% (plasma: 100%; serum: 99.39%; whole blood: 100%) | 94.68% | serum, plasma, whole blood | 10 μl | 10–15 min |
VivaDiagTM COVID-19 IgM / IgG Rapid Test | VivaChek Laboratories, Inc., United States | 100% | Infection time, 4–10 days; IgM and IgG: 81.25% Infection time, 11–24 days; IgM and IgG: 97.1% | Infection time, 4–10 days; IgM and IgG: 94.6% Infection time, 11–24 days; IgM and IgG: 99.3% and 95.1%, respectively | whole blood | 10 μl | 15 min |
SARS-CoV-2 IgG / IgM Rapid Qualitative Test Kit | Xiamen Biotime Biotechnology Co., Ltd., China | – | – | – | – | 10 μl | 10–15 min |
One Step Test for Novel Coronavirus (2019-nCoV) IgM / IgG Antibody | Getein Biotech, Inc., China | 94.1% | 95.1% | – | serum, plasma, fingertip blood, whole blood | 10 μl of serum or plasma; 20 μl of fingertip blood or whole blood | 10–20 min |
Coronavirus Rapid Test COVID-19 IgM / IgG | Zhuhai Encode Medical Engineering Co., Ltd., China | – | – | – | serum, plasma, whole blood | – | – |
Novel Coronavirus (SARS-CoV-2) IgM / IgG Antibody Assay Kit | Shenzhen Watmind Medical Co., Ltd., China | – | – | – | serum, plasma, whole blood, fingertip blood | 10 μl of serum or plasma; 20 μl of fingertip blood or whole blood | 15 min |
SARS-CoV-2 IgM / IgG Antibody Rapid Test | Qingdao Hightop Biotech Co., Ltd., China | IgG: 93%; IgM: 82% | IgG: 97.5%; IgM: 96% | IgG: 96.5%; IgM: 92.8% | serum, plasma, whole blood | 10 μl of serum or plasma; 20 μl of whole blood | 15–20 min |
Cellex qSARS-CoV-2 IgG / IgM Cassette Rapid Test | Cellex, Inc., United States | 98.43% | 96.4% | 97.09% | serum, plasma, whole blood | 10 μl | 15 min |
NADAL COVID-19 IgG / IgM Test | nal von minden GmbH, Germany | 94.1% (86.8%–98.1%) | 99.2% (97.7%–99.8%) | 98.2% (96.6%–99.2%) | serum, plasma, whole blood | 10 μl | 10 min |
Wuhan Coronavirus Rapid Test (2019-nCoV, COVID-19) IgG / IgM | Novazym s.c., Poland | IgG: 91.8%; IgM: 95.7% | IgG: 96.4%; IgM: 97.3% | IgG: 95%; IgM: 96.8% | serum, plasma, whole blood | 5 μl of serum or plasma; 10 μl of whole blood | 15 min |
Abbreviations: IgG, immunoglobulin G; IgM, immunoglobulin M |
Some manufacturers tested the cross-reactivity of their products with antibodies against different pathogens, but none of them tested antibodies against other coronaviruses. In almost all cases, corresponding manuals warn that a negative result does not exclude SARS-CoV-2 infection. They also recommend that results should not be considered as the sole criteria for the diagnosis of COVID-19 and must be interpreted together with other clinical data (and sometimes epidemiological and / or other laboratory results) available to the physician. Some manufactures reported additional limitations and requirements, as listed below:
- ACCU-TELL COVID-19 IgG / IgM Cassette: the continued presence or absence of antibodies may not be any criterion for either success or failure of therapy. Results from immunosuppressed patients should be interpreted carefully.
- 2019-nCoV IgG / IgM Rapid Test Cassette: the hematocrit level needs to be between 25% and 65% to obtain accurate results.
- Rapid test 2019-nCOV whole blood: the hematocrit level between 35% and 65% is recommended for the most accurate results.
- The StrongStep COVID-19 IgG / IgM Combo Test: clinical diagnosis should not be based on the result of a single test, but should be established after considering all the clinical findings, particularly including the reverse transcriptase–polymerase chain reaction test for SARS-CoV-2.
- STANDARD Q COVID-19 IgM / IgG Duo: for more accuracy of patients’ immune status, additional follow-up testing using other laboratory methods is recommended.
- 2019-Novel Coronavirus (2019-nCoV) IgG / IgM GICA Rapid Test Kit: the test results of this product cannot be used as a basis for diagnosis.
- Novel Coronavirus (SARS-CoV-2) IgM / IgG Antibody Assay Kit: hemolysis, lipemia, and microbiological pollution can affect the test result, such specimen is not to be examined. Patients with impaired immunity after immunosuppression therapy, HIV-positive, and / or receiving immunosuppressive treatment after transplant or receiving therapy have impaired immunological response and their results can lead to an incorrect diagnosis. If the infection is suspected, the test should be repeated after 7 to 14 days. Both the first and the second specimen should be examined at the same time to determine whether seroconversion occurred in the primary infection. Each laboratory should work out its own appraisal for their test tubes.
- SARS-CoV-2 IgM / IgG Antibody Rapid Test: color intensity on the T-line may not be associated with the antibodies concentration in the specimen.
- Cellex qSARS-CoV-2 IgG / IgM Cassette Rapid Test: the intensity of the test band does not correlate with the virus titer in the specimen. Viruses with mutations in the epitope recognized by the antibody utilized in the test could provide a negative result. If symptoms persist, along with negative results, it is recommended to resample the patient after a few days or test them with an alternative testing device.
- The NADAL COVID-19 IgG / IgM Test: the continued either presence or absence of antibodies may not be used to determine the success or failure of therapy. The results of immunosuppressed patients should be interpreted with caution. A positive test result can also appear in case of negative polymerase chain reaction results, because antibodies are still present in the blood after the disease and, hence, can be detected.
- Wuhan Coronavirus Rapid Test (2019-nCoV, COVID-19) IgG / IgM: frozen and thawed samples (particularly repeatedly) contain particles that can block the membrane, slow down the flow of reagents, and may lead to a high background color, making the interpretation of results difficult. Specimens containing the unusually high titer of heterophile antibodies or rheumatoid factor also may affect results.
Rapid antibody tests are simple in use, fast, and cheap, but they have several limitations. First of all, they are useless for the detection of infection in early and even mid phases. The number of antibodies produced in response to SARS-CoV-2 infection increases relatively late, several days after the onset of symptoms.2-4 This disqualifies such tests from „on admission” diagnostic workup due to a huge number of false-negative results. This observation was reported in Poland, Czech Republic, Spain, and Italy.5 None of the analyzed kits was tested for potential cross-reactivity with antibodies against other Betacoronaviruses (HKU1, NL63, OC43, or 229E), which could be potentially responsible for false-positive results. The initial World Health Organization protocols and the Wuhan handbook, followed by regional experts, did not recommend the use of rapid antibody tests in the diagnostic procedures related to COVID-19.6 The reference methods for diagnosing patients infected with SARS-CoV-2 involve the molecular approach. Currently, enormous progress in this field is being observed. The Food and Drug Administration approved the use of a few point-of-care molecular analyzers and tests worldwide or limited to the United States territory.7 Also in China and Europe, similar technology has been developed. Bosch Healthcare Solutions presented the rapid molecular diagnostic test on the Vivalytic analysis device.8 However, the shortage of molecular tests, their relatively high prices, particularly as compared with antibody measurements, and the increasing number of convalescents, including persons after an oligosymptomatic and even asymptomatic course of the disease, may change this situation. In particular, the percentage of asymptomatic cases is not clear and ranges from 10% to 80% of the entire infected population.9,10 The molecular testing of a person belonging to this crescent group of convalescents is useless due to the absence of the virus. Therefore, the need for antibody testing was developed by the World Health Organization in the interim guidelines as of March 19, 2020.11 This necessity was also noted by the Food and Drug Administration and other national associations, which approved 5 rapid antibody test kits for detecting SARS-CoV-2 on March 30, 2020.1,12 These tests can be used solely by medical professionals, because only this group can ensure that the samples are processed properly and the results are interpreted correctly. Additionally, the European Commission and the European Union member states are funding fast-track clinical validation studies on rapid diagnostic tests for SARS-CoV-2, conducted by hospital laboratories in several European Union member states. It is expected that the utility and reliability of the subsequent tests will be soon defined.13 It should be clearly stated that, according to current global recommendations, molecular testing is the only way to diagnose the presence of SARS-CoV-2 infection. However, in our opinion, antibody cassette tests for SARS-CoV-2 should be considered in the following situations: 1) to cohort immunocompetent and nonimmunocompetent persons. The cost-effectiveness of such testing is still limited due to the yet relatively low number of convalescents, but within several weeks or months, with an increasing number of ill and convalescent persons, rapid antibody testing would be an easy diagnostic tool of low cost. First of all, it should be recommended for medical staff, soldiers, and persons working close to each other, as well as be used to optimize qualification for quarantine. But finally, population testing should be considered. Healthy, immunocompetent persons may return to normal activities and work, because they have already been infected and, possibly, they are resistant to reinfection and epidemiologically safe (they do not transmit infection); 2) to verify epidemiological monitoring: tracking of patients’ contacts, identification of virus “reservoirs,” the spread of infection in the observed population, and identification of asymptomatic infections in particular; 3) to support epidemiological monitoring of patients in quarantine or isolation in the absence of molecular tests or even as a cheaper alternative for a molecular test. We suggest testing in 7-day intervals (on day 7, 14, and 21). Such a procedure should be efficient for identifying infected patients, even those asymptomatic, and follow-up time is long enough to confirm healthy status in those with negative test results. Also, new technologies could be involved. For example, the patient can take a photo of the cassette after performing the test and share it with the physician, who will professionally interpret the result remotely.
Przemysław Tomasik, MD, PhD, Department of Clinical Biochemistry, Pediatric Institute, Jagiellonian University Medical College, ul. Wielicka 265, 30-663 Kraków, Poland, phone: +48 12 658 06 81, email: p.tomasik@uj.edu.pl
May 29, 2020.
Przemysław Tomasik, Filipina Krótki, Mateusz Jońca, Tomasz Anyszek (PT: Department of Clinical Biochemistry, Pediatric Institute, Jagiellonian University Medical College, Kraków, Poland; FK and MJ: Student’s Research Society affiliated to the Department of Clinical Biochemistry, Pediatric Institute, Jagiellonian University Medical College, Kraków, Poland; TA: Diagnostyka Medical Laboratory, Kraków, Poland)
None declared.
Tomasik P, Krótki F, Jońca M, Anyszek T. Rapid point-of-care antibody cassette tests for severe acute respiratory syndrome coronavirus 2: practical considerations. Pol Arch Intern Med. 2020; 130: 459-462. doi:10.20452/pamw.15311
- 1.
- Flisiak R, Horban A, Jaroszewicz J, et al. Management of SARS-CoV-2 infection: recommendations of the Polish Association of Epidemiologists and Infectiologists as of March 31, 2020. Pol Arch Intern Med. 2020; 130: 352-357.
- 2.
- Zhou P, Yang XL, Wang XG, et al. A pneumonia outbreak associated with a new coronavirus of probable bat origin. Nature. 2020; 579: 270-273.
- 3.
- Shu-Yuan X, Yingjie W, Huan L. Evolving status of the 2019 novel coronavirus infection: Proposal of conventional serologic assays for disease diagnosis and infection monitoring. J Med Virol. 2020; 92: 464-467.
- 4.
- Xiao AT, Gao C, Zhang S. Profile of specific antibodies to SARS-CoV-2: The first report. J Infect. 2020 Mar 21. [Epub ahead of print].
- 5.
- Dibble M. Coronavirus tests Spain purchased from China fail to detect positive cases. Washington Examiner. March 26, 2020. https://www.washingtonexaminer.com/news/coronavirus-tests-spain-purchased-from-china-fail-to-detect-positive-cases. Accessed March 28, 2020.
- 6.
- Liang T, ed. Handbook of COVID-19 Prevention and Treatment. Zhejiang University School of Medicine; 2020.
- 7.
- US Food & Drug Administration. Coronavirus (COVID-19) update: FDA provides more regulatory relief during outbreak, continues to help expedite availability of diagnostics. https://www.fda.gov/news-events/press-announcements/coronavirus-covid-19-update-fda-provides-more-regulatory-relief-during-outbreak-continues-help. Accessed March 28, 2020.
- 8.
- Bosch. Combating the coronavirus pandemic: Bosch develops rapid test for COVID-19. https://www.bosch.com/stories/vivalytic-rapid-test-for-covid-19/. Accessed March 28, 2020.
- 9.
- Nishiura H, Kobayashi T, Suzuki A, et al. Estimation of the asymptomatic ratio of novel coronavirus infections (COVID-19). Int J Infect Dis. 2020 Mar 13. [Epub ahead of print].
- 10.
- Mizumoto K, Kagaya K, Zarebski A, Chowell G. Estimating the asymptomatic proportion of coronavirus disease 2019 (COVID-19) cases on board the Diamond Princess cruise ship, Yokohama, Japan, 2020. Euro Surveill. 2020; 25: 2000180.
- 11.
- World Health Organization. Laboratory testing for coronavirus disease (COVID-19) in suspected human cases: interim guidance, 19 March 2020. https://apps.who.int/iris/handle/10665/331501. Accessed March 28, 2020.
- 12.
- US Food & Drug Administration. FDA approves rapid antibody test kits for COVID-19. https://www.fda.gov.ph/fda-approves-rapid-antibody-test-kits-for-covid-19/. Accessed April 2, 2020.
- 13.
- European Centre for Disease Prevention and Control. An overview of the rapid test situation for COVID-19 diagnosis in the EU/EEA. https://www.ecdc.europa.eu/en/publications-data/overview-rapid-test-situation-covid-19-diagnosis-eueea. Published April 1, 2020. Accessed April 3, 2020.